Description
CBhuailrta-icnte arnisdti hcsousing version.
Pumps available with single or double head.
Standard CE mark.
FM approval optional.
Separate installation possible.
Condensate accumulates when conditioning gas in gas conditioning. It always accumulates when cooling moist sample gas. On one hand this may occur inadvertently if thermal bridges occur in the sample gas lines. On the other hand the deposit of moisture is necessary to protect the measuring cells in the analyser from damage and/ or stabilise measurements. Since the sample gas is often conveyed through the analysis system with suction, the condensate must be pumped off to remove it.
So-called peristaltic pumps are particularly suited for this purpose. They systemically protect the sample gas system from external air and based on the hose material used offer high resistance against the often times highly corrosive condensate. The CPsingle and CPdouble pump series were developed specifically for these severe operating conditions.
CBhuailrta-icnte arnisdti hcsousing version.
Pumps available with single or double head.
Standard CE mark.
FM approval optional.
Separate installation possible.
Flow rate: | 0.3 L/h (50 Hz) / 0.36 L/h (60 Hz) with standard hose |
Inlet vacuum: | max. 0.8 bar |
Inlet pressure: | max. 1 bar |
Protection class: | IP 44 (housing version) IP 40 (built-in version) |
Datasheet |
|
No images found.