Sample gas probe GAS 222.35 Amex

In many applications gas analysis is the key to safe and efficient control of process flows, environmental protection, and quality assurance. In the extractive gas analysis, the location of the gas sampling point is crucial for the reproducibility and accuracy of the analysis results. The specific filter capacity, corrosion resistance, and functional equipment requirements for the probe arise from the composition of the sample gas. However, operating costs are also an important criterion in the selection, as the sampling points are frequently located at hard to access points in the system. Effective particle filter backwashing options and low maintenance characterize the extensive GAS probe series.

Description

  • Heated probe with shut-off valve, upstream filter, weather hood and terminal box.
  • The filter element can easily be removed by turning the handle 90°.
  • The probe body and the area around the screw connection for the heated sample gas line are completely isolated.

  • Heater self-regulating to approx. 120 °C (T3)/70 °C (T4).
  • For dust loads up to 200 g/m³.
  • This probe is suitable for use in explosive areas C-US and CSA approval.

Self-regulating temperature : 120 °C (T3)/70 °C (T4)
Ambient temperature : -20 to +80 °C
Low temperature alarm : Contact open at operating temperature, closes at < 95 °C (T3) resp. < 50 °C (T4);
Electrical data : 115 V-230 V, 50/60 Hz
 Max. operating pressure : 6 bar

Datasheet

No images found.